B-Detection Plus Bauer
B-DETECTION PLUS online gas measuring system for high-pressure applications
The online gas measurement system B-DETECTION PLUS monitors the quality of the compressed air: Measurement of CO, CO2, O2 as well as optional absolute humidity and oil (VOC)1. Automatic, constantly and extremely reliable.
By means of the B-CONTROL control, you can at any time observe compliance with the limit values of the breathing air standard DIN EN 12021: 20141.
In the case of an exceedance of the limit, the control system will shows an alarm on the display by means of an optical warning signal and switch off the system before air contaminated with pollutants enters the breathing air bottles.
An automatic flush valve (optional) ensures that the contaminated air is directed into the open air without interrupting the operation of the system if short-term limit values are exceeded.
With the new B-DETECTION PLUS gas measuring system, you now have a reliable partner at your side, available as version B-DETECTION PLUS i integrated into the compressor, and as stand-alone version B-DETECTION PLUS s. As an operator, you can enjoy the assurance of quality and legal compliance with BAUER's online gas measuring system. B-DETECTION systems can easily be retrofitted to existing BAUER compressors!
MEASUREMENT VALUES | |
---|---|
Carbon dioxide (CO2), carbon monoxide (CO), oxygen (O2), absolute humidity1, VOC1 | |
SYSTEM LIMITATIONS | |
Maximum system pressure | 350 bar |
Maximum permissible charging rate compressor | 850 l/min |
Permissible operating temperature | +5°C... +45°C |
MEASUREMENT METHOD | |
CO2 | Nondispersive infrared spectroscopy (NDIR) |
CO, O2 | electrochemical |
Absolute humidity | Via dew point |
VOC (oil) | Photoionization detector (PID) |
MEASUREMENT VALUES AND MEASUREMENT ACCURACY4 | |
CO2 | 0... 2.000 ppm ± 2.5 % Full scale |
CO | 0 … 25 ppm ± 2 % of measurement range final value |
O2 | 0 % … 40 % O2 ± 1 % of measurement range final value |
Dew-point | -70 ...+60°C / 10 ... 40,000 ppm ± 2°C / 1 ppm + 20% of displayed value |
VOC (oil) | 0 … 1 ppm (based on isobutene as calibration gas) |
FURTHER TECHNICAL SPECIFICATIONS | |
Operating voltage | integrated: 24 VDC; stand-alone: 100-250 VAC, 50/60 Hz |
Flow rate | 1.0 ... 3.0 l/min |
Weight | integrated: 3 kg; stand-alone: 8.5 kg |
Dimensions W × H × D [mm] (with access) |
integrated: 160 x 260 x 92; stand-alone: 462 x 354 x 184 |
SYSTEM TEST/CALIBRATION | |
Visual inspection and leak test | Monthly |
Functions check 2 | Min. 3-monthly |
System check 3 | Annually |
Calibration | CO2 and PID sensor: annually and after replacing sensor; dew point sensor: every 24 months |
1 Optional extra
2 Includes test with test gas; kit with test gas cylinders and pressure reducer (flow meter) available from BAUER KOMPRESSOREN
3 System check includes replacement of electrochemical sensors (CO and O2)
4 Accuracy valid at max. +/- 10 ° C deviation from calibration temperature.